Bacharach Gas Pup Instruções de Operação Página 28

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 58
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 27
OPERATION GA-40 T plus
28
œ SOOT TEST
SOOT NUMBER: 5
ENTER THE SOOT NUMBER
0 to 9 - enter the detected soot number
ENTER - confirm the soot number
If the analyser is set for thre soot tests
,
(5.9.3.7), then you will be prompted to carry out the remaining
measurements. The result that appears is an average of the three values.
The soot number that is shown when the option is left remains stored until a new one is entered or the results
are either printed or stored as a report.
5.6 Graphic
To make it possible to display graphics, the averaged results of the last 100 measurement results (for 10
variables) are stored in the RAM memory. Every single one of the 10 variables which form a data block can
be displayed in a diagram. Data block
,
(Sec. 5.10.2 )
MENU - select the main MENU
 and à - move the cursor to GRAPHIC option
Á - select the option, the following screen might appear
22
24
26
28
30
20
0h
30h
0. Tamb
[°C]
NORMAL
AUTO
[°C]
Tamb
MENU - switches ON/OFF the inserted MENU on the graphic screen
The individual lines of the menu have the following meanings:
Vista de página 27