Bacharach Gas Pup Instruções de Operação

Consulte online ou descarregue Instruções de Operação para Medição, teste e controlo Bacharach Gas Pup. Bacharach Gas Pup Operating instructions Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - GA-40T plus

FLUE GAS ANALYSER GA-40T plus Operating manual 11/2000  E L E C T R O N I C S

Página 2

GENERAL INFORMATION GA-40 T plus 10 3 GENERAL INFORMATION The GA-40 T plusplus is a multi-functional flue gas analyser. Electrochemical sensors are

Página 3

GA-40 T plus CONSTRUCTION 11 4 CONSTRUCTION The picture on page 53 shows a block diagram of the GA-40 T plusplus. Here is a description of the in

Página 4

CONSTRUCTION GA-40 T plus 12 G the gas sensors are working even when the analyser is switched off - so the battery should never be disconnected or

Página 5 - GA-40 T plus

GA-40 T plus CONSTRUCTION 13 Tables 2 and 3 show the technical data for the analogue inputs. Parameter U1, U2 I1, I2 Range -10 V ÷ 10 V -10 mA

Página 6

CONSTRUCTION GA-40 T plus 14 SOCKET Pin No. Signal Description 1 GND common I1(-), U1(-)* U/I1 2 I1+ current I1 (+) 3 U1+ voltage U1

Página 7 - 1 INTRODUCTION

GA-40 T plus CONSTRUCTION 15 Printer keys - PRINT PAPER ON/OFF keys - ON OFF numerals - 0 to 9 . C 4.3.2 Warning lights The warning LED&apos

Página 8 - 2 MAINTENANCE

CONSTRUCTION GA-40 T plus 16 SOCKET Pin No. Signal DESCRIPTION 1 RTS control RS 232 2 RxD Receive Data - RS 232C 3 GND ground 4 TX

Página 9

GA-40 T plus OPERATION 17 5 OPERATION 5.1 USE OF THE KEYBOARD 5.1.1 Description of the keys LIGHT - Switches the backlighting of the display on

Página 10 - 3 GENERAL INFORMATION

OPERATION GA-40 T plus 18 œ PARAMETER Title line 1 FUEL • Option line 2 STORE Command line 3 O2R 11 % Data line 4 FOO

Página 11 - 4 CONSTRUCTION

GA-40 T plus OPERATION 19 0 to 9 . - the alphanumeric keys enter numbers in digital mode or letters in alphabetical mode. Each key has a three

Página 13

OPERATION GA-40 T plus 20 The following screen is shown in the STANDBY mode. ** STANDBY ** madur TIME: 1h 23min OFF C

Página 14 - 4.3.1 Keyboard

GA-40 T plus OPERATION 21  and à - move the cursor to the PARAMETER option Á - select the option. The following screen will appear: œ

Página 15 - 4.3.3 Display

OPERATION GA-40 T plus 22 œ FUELS LIGHT OIL * NATURAL GAS TOWN GAS COKE-OVEN GAS LIQUID GAS BIO-DIESEL EXTRA LIGHT O

Página 16 - 4.4 Power supply

GA-40 T plus OPERATION 23 G Hu - Heating value of the fuel G UNIT - unit in which the fuel is measured œ DEFINE FUEL 1 NAME

Página 17 - 5 OPERATION

OPERATION GA-40 T plus 24 œ BOILER PARAMETER 1 POWER : 22.4 kW 2 CONSUM.: 34.5 m3/h 3 TEMP. : 350 °C G Boiler Power sho

Página 18 - 5.1.4 Editing Text

GA-40 T plus OPERATION 25 œ CURRENT VALUES • VOL. Rel. GAS [ppm] Ÿ 3% O2 CO 438 511 COu 475 624 NO

Página 19 - 5.2.2 Switching on

OPERATION GA-40 T plus 26 ENTER - switches between AVERAGED VALUES and CURRENT VALUES C - the display on the AVERAGED VALUES screen gets "s

Página 20 - 5.2.5 Measurement Parameters

GA-40 T plus OPERATION 27 C - ZEROING the sensor Connect end of the sample hose to the PRESSURE(+) stub. The screen will display the measured pre

Página 21 - Á - choose SELECT FUEL

OPERATION GA-40 T plus 28 œ SOOT TEST SOOT NUMBER: 5 ENTER THE SOOT NUMBER 0 to 9 - enter the detected soot number ENTER - c

Página 22 - À - back to option FUEL

GA-40 T plus OPERATION 29 Number (0..9) and name of the measured value unit on the Y-axis type of graphic (NORMAL or FAST) scale of the Y-axis (

Página 23 -  and à - select min or sec

GA-40 T plus 3 CONTENTS 1 INTRODUCTION ...

Página 24 - Á - call the next screen

OPERATION GA-40 T plus 30 œ FLOW MEASUREMENT 0 m/s C - ZEROING 1 CORR. FACTOR : 0.10 C - Sets the zero point for the m

Página 25 - 5.3.2 Averaged Values

GA-40 T plus OPERATION 31 5.9.1 Device Info  and à - move the cursor to DEVICE INFO Á - select option. œ DEVICE INFO DEVICE...: GA-40Tp

Página 26 - 5.3.3 The Averaging Process

OPERATION GA-40 T plus 32 Take the probe out of the stack !  and à - move the cursor to O2 CELL CALIBRATION Á - select the option. The following

Página 27 -

GA-40 T plus OPERATION 33 ENTER - default setting: Pt 500. The ambient temperature can either be measured with an internal sensor placed in the

Página 28 - 5.6 Graphic

OPERATION GA-40 T plus 34 œ CALIBRATION 1 WITH CO GAS • 2 WITH NO GAS •  and à - move the cursor to the sensor to be ca

Página 29

GA-40 T plus OPERATION 35 œ DEVICE SETTINGS 1 BACKLIGHTING 2 CONTRAST • 3 CLOCK • 4 PRINTER SETTINGS • 5 LANGUAG

Página 30 - 0 m/s

OPERATION GA-40 T plus 36 œ DATE/TIME 12:34 1 TIME : 17.34.54 2 DATE : 23.10.96 4 DATE FORMAT: d-m-y  and à - move th

Página 31 - Á - select option

GA-40 T plus OPERATION 37 NUMBER OF COPIES ENTER - switches between 1/2/3. ADDRESS NO. The instrument can store up to 50 addresses. Number one

Página 32

OPERATION GA-40 T plus 38 5.9.3.7 Soot Test ENTER - select whether 1 or 3 soot tests shall be performed 5.9.3.8 Standard ENTER - sets the instru

Página 33 - Á - select the option

GA-40 T plus OPERATION 39 Explanation of the Table: G Store -type of storage G free - availability of memory (number of free Reports, Banks or Bl

Página 34 - Á - select the gas

GA-40 T plus 4 5.6 Graphic ...

Página 35

OPERATION GA-40 T plus 40 , (5.10.1.3) - edit header  and à - move the cursor to STORE REPORT NO. XX , (5.10.1.4) - if all 30 Reports are occupi

Página 36

GA-40 T plus OPERATION 41 œ REPORT NO.: 12 11:23.52 16.01.94 -------------------- 4 - lines free

Página 37 - 5.9.3.6 Acoustic Warning

OPERATION GA-40 T plus 42 The pictures show screens without text.The screen appears as it was when the report was stored. How the text should be edi

Página 38 - 5.10 Storing of Results

GA-40 T plus OPERATION 43 Starting the continuous storing of results (in Banks) STORE - call the STORE menu check whether enough free memory is a

Página 39 - Á -select option REPORTS

OPERATION GA-40 T plus 44 The last bank can be erased in the BANKS TABLE screen (the Bank may not be active)  and à - move the cursor to the last

Página 40

GA-40 T plus OPERATION 45 5.12 Printer 5.12.1 Printing the contents of a screen This is not possible for the screen (D1). A screen may not be prin

Página 41

OPERATION GA-40 T plus 46 m a d u r GA-40 T plus 19.12.93 13:23.41 ------------------------ FLUE: NATURAL GAS O2R

Página 42 - 5.10.1.4 Erasing Reports

GA-40 T plus BASIC PRINCIPLES OF CALCULATING RESULTS 47 6 BASIC PRINCIPLES OF CALCULATING RESULTS 6.1 Quantities obtained from direct measurement

Página 43 - (5.2.5.2)

BASIC PRINCIPLES OF CALCULATING RESULTS GA-40 T plus 48 λ − excess air number As can be seen, the concentration of "undiluted" CO is the

Página 44 - 5.11 Data-Logger

GA-40 T plus BASIC PRINCIPLES OF CALCULATING RESULTS 49 6.5.2 Mass concentrations relative to the concentration of oxygen in combustion gases As w

Página 45 - 5.12 Printer

GA-40 T plus 5 9.3 Structure of the menu DATA ...

Página 46 - 19.12.93 13:23.41

BASIC PRINCIPLES OF CALCULATING RESULTS GA-40 T plus 50 SL - chimney loss - the percentage of heat produced in combustion process, which is convect

Página 47

GA-40 T plus BASIC PRINCIPLES OF CALCULATING RESULTS 51 The above formula may be transformed with the use of formula Œ into the form: •[ ]λ =−2095

Página 48

BASIC PRINCIPLES OF CALCULATING RESULTS GA-40 T plus 52 O2ref -reference oxygen - the parameter used to calculate relative concentrations of compon

Página 49

GA-40 T plus FRONT VIEW WITH KEYBOARD 53 7 BLOCK DIAGRAM AnalogCPUSupplyTKEYB.Thermistor NTCBATTTERY220 V ACCFILTERSPERISTALTICPUMPPURG. PUMPV

Página 50

BLOCK DIAGRAM GA-40 T plus 54 8 FRONT VIEW WITH KEYBOARD

Página 51 - 6.7 Fuel parameters

GA-40 T plus MENU STRUCTURE 55 9 MENU STRUCTURE 9.1 Structure of the main menu MENUParameter Fuel Choose fuel View/defineAverage timeO2RBoiler

Página 52

MENU STRUCTURE GA-40 T plus 56 9.2 Structure of the menu STORE STOREReports Store Report No.List of Reports View ReportEdit Header EditData Banks A

Página 54 - 8 FRONT VIEW WITH KEYBOARD

 E L E C T R O N I C S Vienna - Austria

Página 56 - [.] [.] [.]

GA-40 T plus INTRODUCTION 7 1 INTRODUCTION Congratulations on your purchase of the microprocessor-controlled flue gas analyser GA-40 T plusplus

Página 57

MAINTENANCE GA-40 T plus 8 2 MAINTENANCE 2.1 Gas sensors The following points are important for maximising the operational life of the cells and a

Página 58

GA-40 T plus MAINTENANCE 9 The GA-40 T plusplus will not normally allow the instrument to be switched off if there are still traces of flue gas in

Comentários a estes Manuais

Sem comentários