Instruction 55-9002 29
Gas Pup
Maintenance
Gas Inlet Filter Replacement
Periodically check the gas inlet filter, located directly behind the gas-inlet
fitting, for signs of dirt. As necessary, replace this filter per Figure 23.
Pump Replacement
The pump may become inoperative due to water or solid particles getting
into the valves. This will require replacement of the pump to correct.
To replace the pump (see Figure 24):
1. Remove the instrument’s bottom plate using a
5
/64" Allen wrench to
remove the plate’s four corner screws. Then remove plate and rubber
bumper to expose the instrument’s interior.
2. Back off the pump bracket screw in the lower corner of the flow block
and rotate bracket arm to free pump.
3. Unplug pump connector from printed circuit board and lift pump
straight out of instrument.
4. Install the new pump by lining up the pump’s inlet and outlet tubes
with the two flow block holes and pushing into place. Swing bracket
arm over pump and tighten—aligning pump as necessary.
5. Leak test pump before use per section Leak Check Basics.
Figure 23. Gas Inlet Filter Replacement
INLET FILTER
(P/N 55-0045)