Gas Pup
22 Instruction 55-9002
Calibration
Calibration
It is recommended that the Gas Pup be calibrated by your nearest
Bacharach Service Center. Calibration, however, can be performed in the
field if your facility has the necessary equipment and qualified personnel
to perform the procedures described in the following sections.
Gas Calibration Equipment Setup
The following equipment is recommended to calibrate the Gas Pup (refer
to the Accessories Section for part numbers):
Calibration Kit (includes gas regulator, tubing, and instrument fitting)
Calibration Gas:
- Zero Air (20.9% Oxygen, balance Nitrogen) (optional)
- 2.5 ±0.05% methane
- 100 ±2% methane
WARNING: If you are not using Bacharach’s calibration kit, be
sure not to apply pressure to the gas inlet from your own
calibration equipment. Use the excess flow calibration method,
otherwise incorrect gas readings may occur.
Connect the components of the calibration kit, calibration gas cylinder,
and Gas Pup together as shown in Figure 18. DO NOT connect the Gas
Inlet Fitting to the Pup until instructed to do so in the calibration procedure.
Figure 18. Calibration Equipment Setup
TUBING
*
CALIBRATION-GAS CYLINDER
2.5% or 100% Methane
(See Accessories Section
for Part Numbers)
*Part of Calibration Kit
GAS INLET FITTING*
(Do not connect
until “AC” is displayed
in READING window)
HI%
LO%
PPM
®
RANGE
NATURAL GAS
READING
ON-DEMAND
GAS REGULATOR
*