Instruction 55-9002 27
Gas Pup
Pressure Sensor Check
The pressure sensor does not require calibration, but it is recommended
that the sensor be checked for accuracy on a periodic basis and to assure
that it is still operational.
To check the pressure sensor, the following equipment is recommended:
• Pressure source of at least 50 psig with an attached adjustable regula-
tor (compressed-air cylinder, company gas, or shop air)
• Pressure measuring device (calibrated gauge, or digital manometer)
• Pressure Hose Kit (P/N 55-0038)
• Assortment of hardware: Tubing, Clamps, and Fittings
Check the pressure sensor as follows:
1. Connect the above equipment as shown in Figure 21.
2. Turn Pup ON and select its high-pressure mode as described under
the previous heading Pressure Operating Mode.
3. Adjust regulator of pressure source for a pressure gauge reading of
50 psig. Observe that the pressure reading on the Pup should be
within ±5% of the reading on the gauge.
4. Shut off regulator and disassemble test equipment.
Figure 21. Pressure Testing the Sensor
PRESSURE
SOURCE
(Compressed Air
Cylinder,
Company Gas,
or Shop Air)
PRESSURE
FITTING
(Do not apply pressure
until “P” is displayed
in READING window)
HI%
LO%
PPM
®
RANGE
NATURAL GA S
READING
PRESSURE HOSE KIT
OPTION (P/N 55-0038)
0
1
0
2
0
3
0
4
0
5
0
6
0
7
0
8
0
9
0
1
0
0
CALIBRATED PRESSURE GAUGE
OR DIGITAL MANOMETER
ADJUSTABLE
REGULATOR
TEE
Use clamps or fitings on all tubing
connections to prevent the
tubing from blowing off when
under pressure.
NOTE:
Calibration