Gas Pup
8 Instruction 55-9002
Turning ON the Gas Pup
Important! To assure that all detectors are properly zeroed, the
Gas Pup must be turned on in a gas-free environment (no
combustibles), and that no pressure is being applied to the
instrument’s pressure port.
It is also recommended that the probe and hose assembly be
connected to the instrument before turning it on.
Refer to Figure 5. The Gas Pup is turned ON by momentarily pressing the
key for 1 second. Listen for the pump to start and observe that all
segments of the display are momentarily turned on. After all segments
have been displayed the word “UEr” (Version) is displayed followed by the
instrument’s software version number. Next, the letter “b” appears in the
RANGE window along with the battery voltage momentarily appearing in
the READING window. The instrument will quickly move to display the
word “GAS”, indicating that the instrument is operating in its gas mode.
After several seconds, the READING window will display a flashing “AS”
(Auto Start). This is the Auto Start portion of the instrument’s operation,
and is used to stabilize the gas sensor in order to establish ‘zero’ gas
readings. Without additional operator intervention, the auto start portion of
the start-up routine will finish, leaving the instrument in its PPM gas range.
Operation