Gas Pup
24 Instruction 55-9002
Calibration
Note: If zero air is being applied at this time, remove it before
proceeding to Step 5.
5. While a steady “AC” is being displayed, apply calibration gas by
simply connecting the tubing from the calibration equipment to the
Pup’s gas-inlet fitting. The on-demand gas regulator will automati-
cally begin to supply the correct amount of calibration gas to the
instrument.
6. As the calibration gas is being applied, “AC” will begin flashing until a
steady gas value is detected for a period of 5 seconds, after which the
display will change to the calibration setting (2.5% for the LO% range,
or 100% for the HI% range). At this time the calibration gas may be
removed.
The instrument’s programming contains safeguards to prevent inaccurate
calibration, or to limit the use of a sensor that may require replacement
due to imminent failure.
Should calibration gas not be applied within the calibration period of
10 seconds, the instrument will drop out of the autocal mode and turn off
(error code 2.3.1 is displayed just before the instrument turns off). The
instrument is turned off to attract the users attention and make it evi-
dent that a proper calibration had not been achieved. The prior calibra-
tion values will be maintained to prevent an unintended change to these
values.