Instruction 24-944820
Initial Setup
3.20 Cal Reminder Period
The analyzer can be set to indicate a calibration reminder during the 60
second warm-up period. Calibration reminders can be preset to occur 6, 8,
10, 12, or 15 months after the last calibration. When the preset period is
exceeded the instrument will display the reminder, and how long since the
sensors were last calibrated. The reminder will be displayed at the end of
the 60 second warm-up period. If a calibration reminder is displayed the
operator can press the RUN/HOLD key to move to the Run/Hold screen for
normal operation. Regular calibration periods of 6 months to a year for all
gas sensors (except Oxygen) are recommended.
Set the calibration reminder period as follows:
1. Display the MAIN MENU by press-
ing the MENU (F2) button. If necessary,
press ESC until MENU appears above
F2 .
2. Use the buttons to highlight SETUP, and
then press ENT to display the SETUP MENU.
3. Use the buttons to highlight CAL RE-
MINDER PERIOD, and then press ENT to
display CAL Reminder Period Menu.
4. Use the buttons to highlight the desired
CAL reminder period, and then press the ENT
key to activate the selection and re-display the
Setup Menu.
Note: The date and time settings must be correct to
get accurate cal reminders.
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1 F2 F3
MENU
SETUP MENU
(more)
Logging
Test ID
Username
Run/Hold Format
Language Selection
Button Sound
CAL Reminder Period
MENU
F1 F2 F3
CAL Reminder Period
6 months
8 months
10 months
12 months
15 months
MENU
F1 F2 F3