Instruction 24-9448 13
Initial Setup
3.14 Battery Charger Selection
When using rechargeable NiMH batteries, the AC
power adapter can be used to charge the batteries
while inside the analyzer. The analyzer’s rapid-
charger circuit, however, must be first turned ON.
IMPORTANT: When using disposable alka-
line batteries, the analyzer’s battery charger
circuit should be OFF to prevent the batter-
ies from overheating.
As a precaution, the charger circuit is automati-
cally toggled to its OFF state when the analyzer is
turned OFF, thus requiring the operator to turn
the charger back ON when needed.
Note: The PCA2 must be left turned ON to
charge the batteries.
Turn the battery charger circuit ON and OFF as
follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight SETUP,
and then press ENT to display the SETUP
MENU.
3. Use the buttons to highlight BATTERY
CHARGER, and then press ENT to display
the BATTERY CHARGER screen.
4. Use the buttons to highlight either OFF
(turn charger OFF) or ON (turn charger ON).
In the example shown, ON has been selected.
5. Press ENT to save the selection and re-display the SETUP MENU.
BATTERY CHARGER
OFF
ON
05/26/06 09:25:30 AM
MENU
F1 F2 F3
MAIN MENU
FUEL
PRESSURE
TEMPERATURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1 F2 F3
MENU
SETUP MENU
(more)
TIME
O2 REFERENCE
PRINT PRESSURE
ZOOM
BATTERY CHARGER
LOGGING
(more)
MENU
F1 F2 F3