Bacharach Flue-Gas Sample Conditioning System Especificações Página 82

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 81
Parts & Service ECA 450
78 Instruction 0024-9400
Symptom
Probable Cause & Remedy
Erratic combustion test screen data.
a. Faulty sensor(s):
- Check that all sensors are
properly installed and their
wiring harnesses are correctly
connected to the analyzer.
Refer to Section 4.4.
- Check sensor calibration per
Section 3.
- Replace any sensor that does
not calibrate, and then
recalibrate.
b. Leak in gas path. Check tightness
of all hose connections on probe, and
tightness of all hose connections
internal to the analyzer.
Sensor does not calibrate properly.
a. Wrong calibration gas or
insufficient flow being applied to
sensor. Ensure that the calibration
setup is correct.
b. Depleted sensor. Replace sensor
and recalibrate.
c. If the NO sensor doesn’t calibrate,
its bias battery may be dead.
Replace bias battery per Section
4.4.2 and recalibrate.
Vista de página 81
1 2 ... 77 78 79 80 81 82 83 84 85 86

Comentários a estes Manuais

Sem comentários