ECA 450 Maintenance
Instruction 0024-9400
65
4.4.1 O
2
, CO, NO, NO
2
, & SO
2
Sensor Replacement
All sensors, with the exception of the HC sensor, are replaced in a similar
manner. Do the following to replace either the O
2
, CO-LO, CO-HI, NO, NO
2
,
or SO
2
sensor.
Material Required:
Medium Phillips screwdriver
Replacement sensor (refer to Section 6.1)
Procedure:
1. Remove two screws from sensor-cover plate, and then remove plate to
gain access to sensor compartment. See Figure 4-3.
2. Unplug the printed circuit board from the sensor being replaced by
carefully pulling the board straight up.
3. Note position of sensor pins; then release sensor from its bayonet-type
holder by turning the sensor counterclockwise. Remove sensor by pulling
it straight up.
Note: Discard an old sensor in accordance with local and federal
hazardous waste disposal laws.
4. Remove replacement sensor from its plastic canister. Note that the CO-
LO, CO-HI, SO
2
, and NO
2
sensors have a shorting-spring installed
between two of their pins. Remove and discard this spring before
installing the sensor.
5. Insert sensor into its holder and turn clockwise to lock in place. Ensure
that the sensor pins are positioned as was noted in Step 3.
6. Plug the circuit board, which was removed in Step 2, into the top of the
sensor.
Important! When replacing the O
2
sensor, be sure that the “+” sensor pin
plugs into the circuit board socket that is also marked “+”. Polarity
markings are located on the bottom of the printed circuit board and the
sensor.
7. Install sensor-cover plate that was removed in Step 1.
8. Allow the sensor that was just installed time to stabilize in the circuit
before continuing with this procedure. Stabilization time for all sensors
(except for the NO sensor) is about 1 hour. An NO sensor requires
approximately 4 hours to stabilize.
9. Place analyzer in an area of fresh air and turn it ON. After its warm-up
cycle completes, observe that no sensor error messages should have
appeared.
10. Calibrate the new sensor per Section 3 (except the O
2
sensor which does
not require calibration).