ECA 450 Operation
Instruction 0024-9400
2.5 Ending a Combustion Test
1. After performing a combustion test per Section 2.4, press RUN key to end
the test. You should hear the pump stop running.
WARNING: Burn hazard! Take all necessary precautions when handling a
hot probe and probe stop.
CAUTION: Allow a hot probe and probe stop to cool before storing them
inside the analyzer’s carrying case.
2. Loosen thumbscrew on probe stop; then remove probe and probe stop
from stack.
Note: If data was saved during the combustion test, the analyzer can be
safely turned off without loosing the saved data, which can be reviewed at
a later time as described in Section “2.9 Recalling Test Data Stored in
Memory.”
3. Turn off analyzer per Section 2.2 Turning Off the Analyzer.
2.6 Performing a Draft Measurement
1. If not already done, turn on analyzer and place probe inside stack as
previously described (Sections 2.1 and 2.4).
2. Press PRESSURE key to display the Pressure screen.
««««««PRESSURE««««««
n.nn xxxx
Where: n.nn = current pressure reading
xxxx = pressure units (refer to Section 2.17.2)
3. If desired, pressure sensor can be zeroed at this time as described below.
To skip the zeroing procedure, proceed to Step 4.
a) With Pressure screen displayed, again press PRESSURE key to
display the Pressure Zero screen.
Pressure Zero
Disconnect Hose
Then Press ENTER
b) Disconnect probe’s pressure hose from analyzer’s PRESSURE
connector.
c) Press ENTER to zero pressure sensor to atmospheric pressure; after
which, reconnect probe hose that was removed in Step 3b.
Reconnect Hose
4. The Pressure screen is now displaying the stack’s draft value as a
negative pressure.