ECA 450 Maintenance
Instruction 0024-9400
61
4 Maintenance
Field maintenance of the ECA 450 is limited to the following:
Battery charging
Water Trap Filter replacement
Printer Paper replacement
Sensor replacement
Cleaning the probe
Resetting the microprocessor
Replacing the internal particulate filter
All other maintenance should be performed by an authorized Bacharach
Service Center (Section 6.3).
4.1 Battery Charging
A fully charged battery pack will power the analyzer for a minimum of
8 hours. When the battery pack is just about depleted, the message “LOW
BATTERY” will appear at the bottom of the screen; after which, the analyzer
will operate from 5 to 15 minutes before automatically shutting itself off.
Recharge the analyzer’s internal battery pack by first plugging the supplied
AC adapter in a source of 100240 VAC, 50/60 Hz power. A green LED on the
adapter should glow, indicating that it is operational. Then insert the
adapter’s output plug into the analyzer’s POWER connector (Figure 1-1).
A fully discharged battery pack should take approximately 11 hours to
recharge. Note that a protection circuit within the analyzer prevents the
battery pack from being over charged.
If the battery pack becomes depleted while the analyzer is being used, or if a
continuous run time of over 8 hours is required, then the AC adapter can be
used to both run the analyzer and charge the battery pack at the same time.