Bacharach Portable Combustion Analyzer Especificações Página 71

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 70
PCA
3 Manual
0024-9472 Rev 3 71
NOTE: If the sensor’s output is too low to be usable, then
the message “Bad Cal
ibration Sensor End of Life, Entry
Not Saved” will appear. The sensor will now be marked as
being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator of calibration fixture and remove the CO cylinder.
7. Attach a combination 1,000 ppm CO and 1,000 ppm H
2
cylinder to the
regulator of the calibration fixture, and then use the ▲▼ and
◄►
buttons to enter an applied value that exactly equals the CO
concentration stamped on the cylinder.
8. Adjust the regulator for a flow meter indication of approximately 2  SCFH,
and then press ENT to display the CALIBRATE H
2
screen.
NOTE: The calibration range is 400 to 1,500ppm.
Calibrating outside this range causes the message “Bad
Calibration Wrong CAL Entry” to appear in the next step.
Figure 6-8. CO
LOW
Sensor Calibration
9. Wait until the measured reading stabilizes (approximately 3  minutes),
and then press ENT to calibrate the H
2
Measured value to that of the
applied value; after which the message “Good Calibration” should briefly
appear followed by the CALIBRATION LIST screen being re-displayed.
NOTE: If the sensor’s output is low, but still usable, then
the message “Good Calibration WARNING Low Sensor
will appear. The sensor will now be marked as being Low
in the DIAGNOSTICS screen.
NOTE: If the sensor’s output is too low to be usable, then
the message “Bad Calibration Sensor End of Life” will
appear followed by the CALIBRATION LIST screen being
re-displayed. The sensor will now be marked as being BAD
in the DIAGNOSTICS screen.
10. Turn OFF the regulator and remove the gas cylinder.
Vista de página 70
1 2 ... 66 67 68 69 70 71 72 73 74 75 76 ... 99 100

Comentários a estes Manuais

Sem comentários